The use of supplemental oxygen plays a vital role in the care of the critically ill preterm infant, but the unrestricted use of oxygen can lead to unintended harms, such as chronic lung disease and retinopathy of prematurity. An overly restricted use of supplemental oxygen may have adverse effects as well. Ideally, continuous monitoring of tissue and cellular oxygen delivery would allow clinicians to better titrate the use of supplemental oxygen, but such monitoring is not currently feasible in the clinical setting. The introduction of pulse oximetry has greatly aided the clinician by providing a relatively easy and continuous estimate of arterial oxygen saturation, but pulse oximetry has several practical, technical, and physiologic limitations. Recent randomized clinical trials comparing different pulse oximetry targets have been conducted to better inform the practice of supplemental oxygen use. This clinical report discusses the benefits and limitations of pulse oximetry for assessing oxygenation, summarizes randomized clinical trials of oxygen saturation targeting, and addresses implications for practice.

The discovery of oxygen is attributed to Polish scientist Michal Sędziwój in 1604, and a series of observations by John Mayow, Carl Wilhelm Scheele, and Joseph Priestley established the necessity of oxygen for life. In the early 1940s, Wilson et al1 demonstrated that the use of 70% oxygen reduced periodic breathing in preterm infants. In 1949, investigators studying breathing irregularities in newborn infants recommended using 40% to 50% oxygen for all preterm infants immediately after birth for as long as 1 month.2 

In 1951, two physicians, Kate Campbell in Melbourne, Australia, and Mary Crosse in Birmingham, England, suggested that unrestricted use of oxygen was associated with an increased risk of retrolental fibroplasia (now called retinopathy of prematurity [ROP]).3,4 Several small clinical studies during the next few years confirmed this suggestion and recommended restricted use of supplemental oxygen.5,9 In those studies, there was a trend toward increased mortality in the oxygen-restricted infants, although it did not reach statistical significance.5,7,9 Therefore, restricted oxygen use in preterm infants gained general acceptance, despite estimates of 16 additional deaths for every case of blindness prevented.10 

Because measurement of arterial oxygen tension was not yet feasible clinically, none of the earlier studies of oxygen supplementation and ROP were able to correlate measures of blood or tissue oxygenation with increased risk of ROP. In 1977, a large, 5-center, prospective observational study could not demonstrate a correlation between high partial pressure of oxygen in arterial blood (Pao2) and ROP but did find a strong association of ROP with cumulative supplemental oxygen exposure.11 In 1987, a small randomized study of transcutaneous oxygen monitoring in infants with a birth weight <1300 g found a significantly lower rate of ROP in infants who were managed with continuous oxygenation measures versus standard intermittent oxygenation assessment.12 

In the ensuing decades, numerous observational studies have indicated that the incidence of ROP and bronchopulmonary dysplasia could be reduced by restricted use of oxygen. In 2007, the Guidelines for Perinatal Care recommended an oxygen saturation range of 85% to 95%.13 Recently completed randomized trials using nearly identical trial designs have now provided additional evidence regarding the effects of varying saturation targets in the NICU. The present clinical report discusses the benefits and limitations of pulse oximetry for assessing oxygenation, summarizes randomized clinical trials of oxygen saturation targeting, and addresses implications for practice.

Pulse oximeters measure the differential absorption of red and infrared light by oxyhemoglobin and deoxyhemoglobin. In neonates and young infants, light is transmitted through a distal extremity and sensed by a detector placed on the opposite side of the extremity. Pulsatile blood flow results in fluctuations in blood volume, thereby changing the distance the light has to travel. Detecting this variable component of light transmission allows pulse oximeters to eliminate signals attributable to nonarterial blood elements, such as venous blood, skin, connective tissue, muscle, and bone, directly measuring the relative amounts of oxyhemoglobin and deoxyhemoglobin in arterial blood and reporting saturation (Spo2).

Device Limitations

Accuracy. The accuracy of pulse oximetry is determined by comparison of Spo2 with the measured saturation of arterial blood (Sao2). Most manufacturers report an SD of the difference between Spo2 and actual Sao2 of 3 points for neonates. However, because 1 SD on each side of the mean includes approximately 68% of the measurements, nearly one-third of the measurements will fall outside that range. For example, an Spo2 reading of 88% could reflect an actual Sao2 between 85% and 91% in 68% of infants but may fall outside a range of 82% to 94% in up to 5% of infants.

The accuracy of pulse oximetry also depends on the range of saturations being measured. Reports of increased inaccuracy at the lower ranges of saturation values commonly encountered in the NICU are of great concern. For oximetry saturation readings in the 85% to 89% range, early studies reported that actual arterial saturations were as much as 10 points lower.14,15 These findings have been confirmed in the most recently developed devices using signal extraction technology to reduce motion artifact; in 1 study, 39% of oximeter readings in the 85% to 89% range had arterial saturations below that range, with 25% of those readings having an actual Sao2 <80%.16 This finding is consistent with a previous observation that using an 85% to 89% Spo2 range resulted in Pao2 values much lower than expected.17 In addition, pulse oximeters are only calibrated down to 80%; saturations below this level are extrapolated and may therefore be subject to even greater error.

Averaging Times. Pulse oximeters do not give instantaneous readings of Spo2 because aberrant signals can make the device response erratic. Modern devices use time-averaging (typically, from 2–16 seconds) over several heartbeats to smooth out the displayed readings. In general, longer averaging times result in a more stable value with fewer false alarms; however, longer averaging times are also less sensitive to brief deviations in saturation outside the targeted range. Longer averaging times not only reduce the detection of desaturations that are either brief (<30 seconds) or marked (<70%) but also overestimate the duration of some detected events by combining 2 or more shorter events.18,19 Shorter averaging time will detect more events but result in more false alarms. Studies have not been able to demonstrate that averaging times alter the amount of time actually spent outside targeted ranges. However, a particular concern is the potential for delayed detection of hypoxemic events.

Pulse Oximeter Algorithms. Pulse oximeters do not measure oxygen saturation directly but derive Spo2 from an internal reference table generated from empirical measurements of Sao2 in healthy adult subjects. No pulse oximeter uses calibration data derived from Sao2 measurements in critically ill patients or even in well infants. Although the effect of age on pulse oximeter accuracy has not been studied, at least 1 study has shown that in critically ill adult patients, changes in Spo2 tend to overestimate actual changes in Sao2, and this discrepancy worsened with decreasing hemoglobin concentrations.20 

Relationship Between Sao2 and Pao2

Oxygen delivery depends on 2 factors: oxygen content of the arterial blood and blood flow. Oxygen content is determined by hemoglobin-oxygen saturation and, to a much lesser extent, by dissolved oxygen; both hemoglobin saturation and dissolved content depend on the prevailing Pao2. Although the relationship between Sao2 and Pao2 is reasonably linear at Sao2 values <80%, the slope of that relationship changes at Sao2 levels >80%, resulting in large changes in Pao2 with small changes in Sao2. This relationship is even more exaggerated in the presence of hemoglobin F, which shifts the oxyhemoglobin dissociation curve to the left. Given that Spo2 is, at best, an estimate of Sao2, Spo2 measurements become poor predictors of actual Pao2 levels, particularly when the infant is receiving supplemental oxygen.

Fetal Versus Adult Hemoglobin

Absent a history of intrauterine transfusion, all extremely low birth weight neonates have high concentrations (>95%) of hemoglobin F in their blood. Hemoglobin F has a higher affinity for oxygen than does hemoglobin A and enhances tissue oxygen delivery at lower Sao2 levels. As the amount of hemoglobin A relative to hemoglobin F increases in the blood (eg, after a red blood cell transfusion), this ability diminishes. Because the absorption spectrum for hemoglobin F is similar to hemoglobin A, there is no effect on the correlation between Spo2 and Sao2.

Few studies have examined ways to best target a specific oxygen saturation range in preterm infants. Manually maintaining oxygen saturation targets in a given range depends on several factors, including: (1) technology (ie, setting Spo2 alarm limits); (2) personnel (bedside nurses); and (3) the clinical stability of the patient. Although automated, closed-loop systems of oxygen delivery have been developed, they are not approved for clinical use in the United States.21 

Alarm Limits

Alarm limits must be distinguished from targets. Targets represent the clinical goal, and alarm limits are used to achieve that goal. In clinical practice, alarm limits typically are set at or slightly beyond the target range. Some monitoring systems allow the use of “alerts” or “soft” alarms, which are less disruptive (being either visual, or at a lower volume or frequency) but warn that a parameter is about to reach an alarm limit. In these cases, the alerts are set within the targets, and the alarm limits may be set wider.

From a human engineering perspective, there are 2 problems with the setting of alarm limits. First, the majority of alarms do not require intervention. Most are either false (eg, a displaced probe or electrode) or are so brief that an intervention is not required. Second, the sheer number of alarms that go off in a busy NICU in a single day can total in the thousands, leading to desensitization. Both issues can lead to disregard of alarms, either deliberately or unintentionally; this condition has been termed “alarm fatigue” and is one reason why providers change alarm limits from those ordered. Clucas et al22 observed that in infants weighing <1500 g, the lower alarm limit was set correctly 91% of the time, but the upper alarm limit was set correctly only 23% of the time. This differential compliance with low versus high alarms could be attributable to an increased tendency for the high alarm limit to be reached, the assumption that hypoxemia is more detrimental than hyperoxemia, and/or the fact that many monitors automatically reset to a high alarm limit of 100% when first turned on.23 

A balance must be struck between setting alarm limits too narrow (increasing the number of unnecessary alarms) or too wide (decreasing the safety margin for intervention). Studies have shown that matching the alarm limits with the target range is associated with more time spent within the target range.24,25 

Personnel

In the multicenter COT (Canadian Oxygen Trial), study participants were maintained within the intended Spo2 range between 68% and 79% of the time. Nurses from one of the centers identified several factors as important in targeting a specific saturation range, including: (1) education; (2) prompt response times; and (3) a favorable nurse-to-patient ratio.26 Targets in the Canadian trial were achieved significantly more often than in other randomized studies,25,27 even though those studies also used educational interventions and process algorithms.24,28 Even in studies in which favorable nurse-to-patient ratios were believed to exist, infants spent 33% to 38% of the time outside their target ranges.20,25 Maintaining infants in a given target range is an extremely labor-intensive process, as evidenced by studies showing that multiple manual adjustments per hour only achieved target ranges approximately 50% of the time.29 Using a fully automated oxygen-controlling system improved targeting by 7% over manual control.30 

An additional concern is that manual documentation of hyperoxemic and hypoxemic episodes results in significant underreporting of such events.31,32 Better tracking of saturation targeting can be accomplished by using third-party data extraction technology33 or by using the histogram feature available on some monitoring equipment.27,34 

Stability of the Saturation Signal in Clinical Settings

Preterm infants who require respiratory support are at increased risk of straying outside desired oxygen saturation targets, particularly if they are receiving supplemental oxygen. Because these infants often have desaturations during routine care (eg, repositioning, feeding, suctioning), it was once common practice to increase supplemental oxygen just before delivering such care (ie, preoxygenation). Preoxygenation also has been used commonly during intubation or other invasive procedures. Such practices may be harmful.35 Instead, oxygen saturation values should be monitored closely, with measures to increase oxygenation used only as needed to maintain Spo2 within the target range.

The optimal saturation range for preterm infants in the NICU has remained elusive for more than 70 years. Although studies performed more than 50 years ago suggested an increased mortality associated with restricted oxygen administration,36 observational trials performed in the era of continuous Spo2 monitoring suggest that mortality is unchanged, with target Spo2 ranges as low as 70%.37 In addition, data from the Vermont Oxford Network indicate that the incidences of ROP and bronchopulmonary dysplasia are lower when a lower oxygen saturation range is targeted.38 However, because these were observational studies, no cause-and-effect relationship can be inferred.

The first published randomized controlled trial (RCT) of differential targeting of oxygen saturations was the STOP-ROP (Supplemental Therapeutic Oxygen for Prethreshold Retinopathy of Prematurity) trial, published in 2000.39 This study randomized infants to treatment when they reached “prethreshold” ROP, at an average postnatal age of 10 weeks. In this multicenter trial, 649 infants with prethreshold ROP were randomized to a saturation range of 89% to 94% (conventional arm) or 96% to 99% (supplemental arm). Progression to threshold ROP was not significantly different between groups in the total population; however, significant benefit was observed for infants in the high oxygen saturation arm who did not have “plus disease” (abnormal dilation and tortuosity of posterior pole blood vessels). On the negative side, infants in the high-oxygen saturation arm experienced an increased length of supplemental oxygen therapy and more often received diuretics at 50 weeks’ postmenstrual age.

A second RCT that randomized infants to treatment at a later postnatal age was the BOOST (Benefits of Oxygen Saturation Targeting) trial (N = 358 infants), which hypothesized that maintaining higher oxygen saturation target ranges (95%–98% vs 93%–96%) would improve growth and neurodevelopmental outcomes.40 The pulse oximeters in both groups were modified to read a targeted value in the range of 93% to 96%. The study reported no benefit to the higher saturation range but did find, similar to the STOP-ROP trial, that infants in the high-saturation arm had significant increases in length of oxygen therapy, supplemental oxygen at 36 weeks’ corrected gestation, and home oxygen.

In 2003, an international meeting of clinical trials experts, statisticians, neonatologists, ophthalmologists, and developmental pediatricians was convened to harmonize the planned RCTs of different target saturation ranges to be able to conduct a prospective individual patient meta-analysis of the data after completion of the follow-up phase of the individual trials (NeOProM [Neonatal Oxygenation Prospective Meta-analysis]).41 Investigators from all 3 planned studies agreed, including SUPPORT (Surfactant Positive Airway Pressure and Pulse Oximetry Trial), sponsored by the Eunice Kennedy Shriver National Institute for Child Health and Human Development; the BOOST-II United Kingdom, Australia, and New Zealand study groups; and the COT trial. Although there were small differences in study design and outcome measures (Table 1), the studies were similar in terms of the population enrolled, methods, interventions tested, and outcomes collected. All studies were masked by the use of pulse oximeters that read 3% above or below the infant’s actual saturation value within the 85% to 95% range. Outside the range of study saturation values (≤84% and ≥96%), true saturation values were displayed. The primary outcome of the NeOProM study was a composite of death or disability at 18 to 24 months of corrected age. It was estimated that 5000 infants would be needed to detect a 4% difference in the rate of death or disability.42 

TABLE 1

RCTs of Differing Pulse Oximetry Targets

StudyPrimary OutcomePrimary Outcome ResultsOther Findings
STOP-ROP39  Rate of progression to threshold ROP (89%–94% vs 96%–99%) No significant differences • Higher saturation range exhibited worsening of chronic lung disease and longer duration of hospitalization 
N = 649 
BOOST40  Growth and developmental outcomes (91%–94% vs 95%–98%) No significant differences • Higher saturation range required oxygen for a longer period of time, dependence on oxygen at 36 wk postmenstrual age, and need for home oxygen 
N = 358 
SUPPORT43,44  Death, severe ROP, or both (85%–89% vs 91%–95%) No significant differences • Severe ROP significantly more common in the higher Sao2 range 
N = 1316 • Increased mortality in the lower Sao2 range at 18–22 mo of corrected age 
• No significant difference in the composite outcome of death or neurodevelopmental impairment at 18–22 mo 
BOOST II45,48  Death or neurodevelopmental impairment at 18–22 mo of corrected for prematurity (85%–89% vs 91%–95%) No significant differences in a pooled analysis of all 3 trials47  • Change in oximeter algorithm during the study 
No significant difference in individual trial analyses46,48  • Study stopped before complete enrollment 
N = 2448 In a post hoc analysis combining 2 of the 3 trials, the primary outcome occurred in 492 (48.1%) of 1022 in the lower target group versus 437 (43.1%) of 1013 in the higher target group (RR, 1.11 [95% CI, 1.01–1.23]; P = .023)46  • Severe ROP significantly more common in the higher Sao2 range 
• Significantly increased necrotizing enterocolitis at the lower saturation range 
• Significantly increased mortality at hospital discharge in the lower Sao2 range with the revised oximeter algorithm 
COT49  Death before a corrected age of 18 mo or survival with ≥1 of the following: gross motor disability, cognitive or language delay, severe hearing loss, and bilateral blindness (85%–89% vs 91%–95%) No significant differences • Change in oximeter algorithm during the study 
 N = 1201  • No difference in mortality 
   • Targeting the lower saturation range reduced the postmenstrual age at last use of oxygen therapy 
StudyPrimary OutcomePrimary Outcome ResultsOther Findings
STOP-ROP39  Rate of progression to threshold ROP (89%–94% vs 96%–99%) No significant differences • Higher saturation range exhibited worsening of chronic lung disease and longer duration of hospitalization 
N = 649 
BOOST40  Growth and developmental outcomes (91%–94% vs 95%–98%) No significant differences • Higher saturation range required oxygen for a longer period of time, dependence on oxygen at 36 wk postmenstrual age, and need for home oxygen 
N = 358 
SUPPORT43,44  Death, severe ROP, or both (85%–89% vs 91%–95%) No significant differences • Severe ROP significantly more common in the higher Sao2 range 
N = 1316 • Increased mortality in the lower Sao2 range at 18–22 mo of corrected age 
• No significant difference in the composite outcome of death or neurodevelopmental impairment at 18–22 mo 
BOOST II45,48  Death or neurodevelopmental impairment at 18–22 mo of corrected for prematurity (85%–89% vs 91%–95%) No significant differences in a pooled analysis of all 3 trials47  • Change in oximeter algorithm during the study 
No significant difference in individual trial analyses46,48  • Study stopped before complete enrollment 
N = 2448 In a post hoc analysis combining 2 of the 3 trials, the primary outcome occurred in 492 (48.1%) of 1022 in the lower target group versus 437 (43.1%) of 1013 in the higher target group (RR, 1.11 [95% CI, 1.01–1.23]; P = .023)46  • Severe ROP significantly more common in the higher Sao2 range 
• Significantly increased necrotizing enterocolitis at the lower saturation range 
• Significantly increased mortality at hospital discharge in the lower Sao2 range with the revised oximeter algorithm 
COT49  Death before a corrected age of 18 mo or survival with ≥1 of the following: gross motor disability, cognitive or language delay, severe hearing loss, and bilateral blindness (85%–89% vs 91%–95%) No significant differences • Change in oximeter algorithm during the study 
 N = 1201  • No difference in mortality 
   • Targeting the lower saturation range reduced the postmenstrual age at last use of oxygen therapy 

COT, Canadian Oxygen Trial; BOOST, Benefits of Oxygen Saturation Targeting; STOP, Supplemental Therapeutic Oxygen for Prethreshold Retinopathy of Prematurity; SUPPORT, Surfactant Positive Airway Pressure and Pulse Oximetry Trial.

The first of these 3 RCTs to be published was SUPPORT.43 In this study, infants between 240/7 weeks’ and 276/7 weeks’ gestational age (N = 1316) were randomized to the 2 different oxygen saturation ranges (85%–89% or 91%–95%) and also to either CPAP or intubation and surfactant, in a factorial design. Oxygen saturation targeting was initiated within 2 hours of birth. The primary outcome was a composite of severe ROP (defined as the presence of threshold retinopathy, need for surgical intervention, or the use of bevacizumab), death before discharge from the hospital, or both. The oximeters in SUPPORT used an older software algorithm that subsequently was updated for the other RCTs.

The composite primary outcome in SUPPORT did not differ significantly between the lower and the higher oxygen saturation groups (28.3% vs 32.1%; relative risk [RR], 0.90; 95% confidence interval [CI], 0.76–1.06). However, death before discharge from the NICU was significantly different, occurring in 19.9% of infants in the lower oxygen saturation group and 16.2% of infants in the higher oxygen saturation group (RR, 1.27; 95% CI, 1.01–1.60), with a number-needed-to-harm of 27. In contrast, the rate of severe ROP among survivors was 8.6% in the lower saturation group versus 17.9% in the higher saturation group (RR, 0.52 [95% CI, 0.37–0.73]), with a number-needed-to-benefit of 11. At 18 to 22 months of corrected age, death or neurodevelopmental impairment occurred in 30.2% of infants in the lower oxygen saturation group and 27.5% of those in the higher oxygen saturation group (RR, 1.12 [95% CI, 0.94–1.32]).44 Mortality remained significantly higher in the lower oxygen saturation group (22.1% vs 18.2%; RR, 1.25 [95% CI, 1.00–1.25]). No significant differences were detected in neurodevelopmental impairment, cerebral palsy, or blindness.

The next RCT published was BOOST-II, from the United Kingdom, Australia, and New Zealand.45 Oxygen saturation targeting began in the first 24 hours of life but not as early as in the SUPPORT study. During these trials, investigators in the United Kingdom found that the standard oximeters (Masimo Corporation, Irvine, California) returned an unexpectedly low number of oxygen saturation values between 87% and 90%. They discovered that there was a shift-up in the oximeter calibration curve that caused values between 87% and 90% to read 1% to 2% higher. A new software algorithm was expected to improve oxygen saturation targeting, although that was not tested. The United Kingdom and Australian investigators began using oximeters with the new software approximately halfway through the trial. However, the New Zealand trial oximeters were not modified, because enrollment had already been completed. Of 2448 infants enrolled in BOOST-II, 1187 (48.5%) were monitored with oximeters incorporating the new software.

Because of the increased mortality in the lower oxygen saturation range in SUPPORT, the BOOST-II Data Safety and Monitoring Board conducted a safety analysis in December 2010.50 In the 1187 infants monitored with the revised algorithm, those assigned to the lower target range had a significantly increased mortality rate at 36 weeks’ gestational age (23.1% vs 15.9%; RR, 1.45 [95% CI, 1.15–1.84]). However, among the entire study population (N = 2448), there was no significant difference. The rate of ROP requiring treatment was reduced in the lower saturation group (10.6% vs 13.5%; RR, 0.79 [95% CI, 0.63–1.00]), and the rate of necrotizing enterocolitis requiring surgery or causing death was increased in that group (10.4% vs 8.0%; RR, 1.32 [95% CI, 1.02–1.68]). The rate of bronchopulmonary dysplasia was unaffected. Although a recent report combining outcomes for 2 of the 3 BOOST-II sites found a significant difference in the composite outcome of death or disability by 2 years of age in a post hoc analysis,46 a pooled analysis from all 3 BOOST-II sites, as originally planned, showed no significant difference in this outcome between the 2 arms (46.8% in the lower vs 43.4% in the higher saturation group; P = .10).47 

Two-year outcomes for the COT were published.49 The primary outcome measure for this study was the rate of death (before 18 months of age) or survival with 1 or more disabilities (gross motor disability, severe hearing loss, bilateral blindness, and cognitive or language delay). Infants were randomly assigned to the lower saturation group or higher saturation group in the first 24 hours of life. Similar to BOOST-II, the calibration software for the oximeter was changed at the midpoint in the study. The number of infants enrolled was 1201, of whom 538 were monitored with oximeters using the new software. There was no difference in the primary composite outcome (51.6% in the lower vs 49.7% in the higher saturation range). Mortality was 16.6% in the 85% to 89% group and 15.3% in the 91% to 95% group. Infants in the lower saturation group had a shorter duration of supplemental oxygen but no changes in any other outcomes. Use of the revised oximeter software had no effect on the primary outcome or mortality.

Saugstad and Aune51 published a systematic review of the 5 oxygen saturation trials. In total, 4911 infants were enrolled in the studies. At the time of this meta-analysis (in 2014), the composite outcome of death or severe neurosensory disability at 18 to 24 months of age was only available for SUPPORT and COT, and there was no difference in that composite outcome between groups. The RR of mortality using the original software in the BOOST-II and COT trials was 1.04 (95% CI, 0.88–1.22). With the revised software (COT and BOOST-II United Kingdom and Australia), the RR of mortality in the lower saturation arm was 1.41 (95% CI, 1.14–1.74). For all 5 trials (SUPPORT; BOOST-II United Kingdom, Australia, and New Zealand; and COT), the risk of mortality was increased (RR, 1.18 [95% CI, 1.04–1.34]). Severe ROP was significantly reduced in the low saturation group (RR, 0.74 [95% CI, 0.59–0.92]), and the risk of necrotizing enterocolitis was increased (RR, 1.25 [95% CI, 1.05–1.49]). The rates of bronchopulmonary dysplasia, patent ductus arteriosus, and intraventricular hemorrhage grades 2 through 4 were not significantly different.

A more recent systematic review52 of the 5 oxygen saturation trials concluded that although infants randomly assigned to the more liberal oxygen target ranges had higher survival rates (relative effect, 1.18 [95% CI, 1.03–1.36]) to discharge, the quality of evidence (assessed by using the Grading of Recommendations Assessment, Development and Evaluation approach53) for this estimate of effect was low for 1 or more of the following reasons: (1) the pulse oximeter algorithm was modified partway into the study); (2) the distribution of Spo2 values did not achieve the planned degree of separation (the median Spo2 in the 85% to 89% groups was >90%); (3) the BOOST-II trials were stopped prematurely on the basis of this outcome; and (4) the COT trial did not report on this outcome explicitly. In addition, although the investigators noted that necrotizing enterocolitis occurred less frequently in the higher saturation arms, there were no significant differences in bronchopulmonary dysplasia, ROP, hearing loss, or death or disability at 24 months of age.52 

The mechanism(s) by which maintaining lower oxygen saturation levels might increase the risk of death is unclear, as the data from these trials suggest that tissue hypoxia was unlikely to be a factor.23 In particular, in the SUPPORT trial, the proportion of infants with median oxygen saturations <85% was no different between the low and high saturation groups.43 Conversely, a post hoc analysis from the SUPPORT trial found a disproportionally higher mortality rate in small-for-gestational-age infants in the lower oxygen saturation target group, suggesting a possible interaction54; if this observation can be confirmed in the other oxygen saturation trials, and more importantly in the individual patient analysis, it would suggest that small-for-gestational-age infants may be more vulnerable to lower oxygen saturations.

In the 5 RCTs discussed in this report, the degree to which individual infants may have been harmed or benefited by the oxygen saturation targets to which they were assigned is not clear.55 Specifically, it would be helpful to know whether an individual infant’s outcome correlated with the amount of time he or she spent within, above, or below the target oxygen saturation range. This information is particularly relevant to ROP because avoiding hypoxemic episodes may be as important as avoiding hyperoxemic episodes.56,59 The preplanned individual patient meta-analysis of these trials (NeOProM) may shed some light on these critical questions.

Establishing a target range for oxygen saturation in infants of extremely low birth weight has both clinical and practical considerations, and the ideal target range remains an elusive goal. Nevertheless, data from several well-designed RCTs can inform practice. Pending additional data, including the individual patient meta-analysis (NeOProM), the following can be concluded:

  1. The ideal physiologic target range for oxygen saturation for infants of extremely low birth weight is likely patient-specific and dynamic and depends on various factors, including gestational age, chronologic age, underlying disease, and transfusion status.

  2. The ideal physiologic target range is a compromise among negative outcomes associated with either hyperoxemia (eg, ROP, bronchopulmonary dysplasia) or hypoxemia (eg, necrotizing enterocolitis, cerebral palsy, death). Recent RCTs suggest that a targeted oxygen saturation range of 90% to 95% may be safer than 85% to 89%, at least for some infants. However, the ideal oxygen saturation range for extremely low birth weight infants remains unknown.

  3. Alarm limits are used to avoid potentially harmful extremes of hyperoxemia or hypoxemia. Given the limitations of pulse oximetry and the uncertainty that remains regarding the ideal oxygen saturation target range for infants of extremely low birth weight, these alarm limits could be fairly wide. Regardless of the chosen target, an upper alarm limit approximately 95% while the infant remains on supplemental oxygen is reasonable. A lower alarm limit will generally need to extend somewhat below the lower target, as it must take into account practical and clinical considerations, as well as the steepness of the oxygen saturation curve at lower saturations.

CI

confidence interval

Pao2

partial pressure of oxygen in arterial blood

ROP

retinopathy of prematurity

RR

relative risk

Sao2

measured saturation of arterial blood

Spo2

pulse oxygen saturation

This document is copyrighted and is property of the American Academy of Pediatrics and its Board of Directors. All authors have filed conflict of interest statements with the American Academy of Pediatrics. Any conflicts have been resolved through a process approved by the Board of Directors. The American Academy of Pediatrics has neither solicited nor accepted any commercial involvement in the development of the content of this publication.

Clinical reports from the American Academy of Pediatrics benefit from expertise and resources of liaisons and internal (AAP) and external reviewers. However, clinical reports from the American Academy of Pediatrics may not reflect the views of the liaisons or the organizations or government agencies that they represent.

The guidance in this report does not indicate an exclusive course of treatment or serve as a standard of medical care. Variations, taking into account individual circumstances, may be appropriate.

All clinical reports from the American Academy of Pediatrics automatically expire 5 years after publication unless reaffirmed, revised, or retired at or before that time.

FUNDING: No external funding.

POTENTIAL CONFLICTS OF INTEREST: Dr Cummings is a consultant for ONY, Inc and Windtree Therapeutics (formerly Discovery Laboratories). Dr Polin is a consultant for Windtree Therapeutics and Fisher & Paykel.

1
Wilson
J
,
Long
S
,
Howard
P
.
Respiration of premature infants: response to variations of oxygen and to increased carbon dioxide in inspired air.
Am J Dis Child
.
1942
;
63
(
6
):
1080
1085
2
Howard
PJ
,
Bauer
AR
.
Irregularities of breathing in the newborn period.
Am J Dis Child
.
1949
;
77
(
5
):
592
609
[PubMed]
3
Campbell
K
.
Intensive oxygen therapy as a possible cause of retrolental fibroplasia; a clinical approach.
Med J Aust
.
1951
;
2
(
2
):
48
50
[PubMed]
4
Crosse
V
.
The problem of retrolental fibroplasia in the city of Birmingham.
Trans Ophthalmol Soc U K
.
1951
;
71
:
609
612
5
Lanman
JT
,
Guy
LP
,
Dancis
J
.
Retrolental fibroplasia and oxygen therapy.
J Am Med Assoc
.
1954
;
155
(
3
):
223
226
[PubMed]
6
Engle
MA
,
Baker
DH
,
Baras
I
,
Freemond
A
,
Laupus
WE
,
Norton
EW
.
Oxygen administration and retrolental fibroplasia.
AMA Am J Dis Child
.
1955
;
89
(
4
):
399
413
[PubMed]
7
Kinsey
VE
.
Retrolental fibroplasia; cooperative study of retrolental fibroplasia and the use of oxygen.
AMA Arch Opthalmol
.
1956
;
56
(
4
):
481
543
[PubMed]
8
Patz
A
,
Hoeck
LE
,
De La Cruz
E
.
Studies on the effect of high oxygen administration in retrolental fibroplasia. I. Nursery observations.
Am J Ophthalmol
.
1952
;
35
(
9
):
1248
1253
[PubMed]
9
Weintraub
DH
,
Tabankin
A
.
Relationship of retrolental fibroplasia to oxygen concentration.
J Pediatr
.
1956
;
49
(
1
):
75
79
[PubMed]
10
Bolton
DP
,
Cross
KW
.
Further observations on cost of preventing retrolental fibroplasia.
Lancet
.
1974
;
1
(
7855
):
445
448
[PubMed]
11
Kinsey
VE
,
Arnold
HJ
,
Kalina
RE
, et al
.
PaO2 levels and retrolental fibroplasia: a report of the cooperative study.
Pediatrics
.
1977
;
60
(
5
):
655
668
[PubMed]
12
Bancalari
E
,
Flynn
J
,
Goldberg
RN
, et al
.
Transcutaneous oxygen monitoring and retinopathy of prematurity.
Adv Exp Med Biol
.
1987
;
220
:
109
113
[PubMed]
13
American Academy of Pediatrics and the American College of Obstetricians and Gynecologists
.
Guidelines for Perinatal Care
. 6th ed.
Elk Grove Village, IL
:
American Academy of Pediatrics
;
2007
14
Brockway
J
,
Hay
WW
 Jr
.
Prediction of arterial partial pressure of oxygen with pulse oxygen saturation measurements.
J Pediatr
.
1998
;
133
(
1
):
63
66
[PubMed]
15
Workie
FA
,
Rais-Bahrami
K
,
Short
BL
.
Clinical use of new-generation pulse oximeters in the neonatal intensive care unit.
Am J Perinatol
.
2005
;
22
(
7
):
357
360
[PubMed]
16
Rosychuk
RJ
,
Hudson-Mason
A
,
Eklund
D
,
Lacaze-Masmonteil
T
.
Discrepancies between arterial oxygen saturation and functional oxygen saturation measured with pulse oximetry in very preterm infants.
Neonatology
.
2012
;
101
(
1
):
14
19
[PubMed]
17
Quine
D
,
Stenson
BJ
.
Arterial oxygen tension (Pao2) values in infants <29 weeks of gestation at currently targeted saturations.
Arch Dis Child Fetal Neonatal Ed
.
2009
;
94
(
1
):
F51
F53
[PubMed]
18
Ahmed
SJ
,
Rich
W
,
Finer
NN
.
The effect of averaging time on oximetry values in the premature infant.
Pediatrics
.
2010
;
125
(
1
). Available at: www.pediatrics.org/cgi/content/full/125/1/e115
[PubMed]
19
Vagedes
J
,
Poets
CF
,
Dietz
K
.
Averaging time, desaturation level, duration and extent.
Arch Dis Child Fetal Neonatal Ed
.
2013
;
98
(
3
):
F265
F266
[PubMed]
20
Perkins
GD
,
McAuley
DF
,
Giles
S
,
Routledge
H
,
Gao
F
.
Do changes in pulse oximeter oxygen saturation predict equivalent changes in arterial oxygen saturation?
Crit Care
.
2003
;
7
(
4
):
R67
[PubMed]
21
Claure
N
,
Bancalari
E
.
Automated closed loop control of inspired oxygen concentration.
Respir Care
.
2013
;
58
(
1
):
151
161
[PubMed]
22
Clucas
L
,
Doyle
LW
,
Dawson
J
,
Donath
S
,
Davis
PG
.
Compliance with alarm limits for pulse oximetry in very preterm infants.
Pediatrics
.
2007
;
119
(
6
):
1056
1060
[PubMed]
23
Sola
A
,
Golombek
SG
,
Montes Bueno
MT
, et al
.
Safe oxygen saturation targeting and monitoring in preterm infants: can we avoid hypoxia and hyperoxia?
Acta Paediatr
.
2014
;
103
(
10
):
1009
1018
[PubMed]
24
Clarke
A
,
Yeomans
E
,
Elsayed
K
, et al
.
A randomised crossover trial of clinical algorithm for oxygen saturation targeting in preterm infants with frequent desaturation episodes.
Neonatology
.
2015
;
107
(
2
):
130
136
[PubMed]
25
Hagadorn
JI
,
Furey
AM
,
Nghiem
TH
, et al;
AVIOx Study Group
.
Achieved versus intended pulse oximeter saturation in infants born less than 28 weeks’ gestation: the AVIOx study.
Pediatrics
.
2006
;
118
(
4
):
1574
1582
[PubMed]
26
Armbruster
J
,
Schmidt
B
,
Poets
CF
,
Bassler
D
.
Nurses’ compliance with alarm limits for pulse oximetry: qualitative study.
J Perinatol
.
2010
;
30
(
8
):
531
534
[PubMed]
27
Lim
K
,
Wheeler
KI
,
Gale
TJ
, et al
Oxygen saturation targeting in preterm infants receiving continuous positive airway pressure. J Pediatr.
2014
;164(4):730–736.e1
28
Ford
SP
,
Leick-Rude
MK
,
Meinert
KA
, et al
.
Overcoming barriers to oxygen saturation targeting.
Pediatrics
.
2006
;
118
(
suppl 2
):
S177
S186
[PubMed]
29
van der Eijk
AC
,
Dankelman
J
,
Schutte
S
,
Simonsz
HJ
,
Smit
BJ
.
An observational study to quantify manual adjustments of the inspired oxygen fraction in extremely low birth weight infants.
Acta Paediatr
.
2012
;
101
(
3
):
e97
e104
[PubMed]
30
Waitz
M
,
Schmid
MB
,
Fuchs
H
,
Mendler
MR
,
Dreyhaupt
J
,
Hummler
HD
. Effects of automated adjustment of the inspired oxygen on fluctuations of arterial and regional cerebral tissue oxygenation in preterm infants with frequent desaturations. J Pediatr.
2015
;166(2):240–244.e1
31
Brockmann
PE
,
Wiechers
C
,
Pantalitschka
T
,
Diebold
J
,
Vagedes
J
,
Poets
CF
.
Under-recognition of alarms in a neonatal intensive care unit.
Arch Dis Child Fetal Neonatal Ed
.
2013
;
98
(
6
):
F524
F527
[PubMed]
32
Ruiz
TL
,
Trzaski
JM
,
Sink
DW
,
Hagadorn
JI
.
Transcribed oxygen saturation vs oximeter recordings in very low birth weight infants.
J Perinatol
.
2014
;
34
(
2
):
130
135
[PubMed]
33
Cirelli
J
,
McGregor
C
,
Graydon
B
,
James
A
.
Analysis of continuous oxygen saturation data for accurate representation of retinal exposure to oxygen in the preterm infant.
Stud Health Technol Inform
.
2013
;
183
:
126
131
[PubMed]
34
Bizzarro
MJ
,
Li
FY
,
Katz
K
,
Shabanova
V
,
Ehrenkranz
RA
,
Bhandari
V
.
Temporal quantification of oxygen saturation ranges: an effort to reduce hyperoxia in the neonatal intensive care unit.
J Perinatol
.
2014
;
34
(
1
):
33
38
[PubMed]
35
Sola
A
,
Saldeño
YP
,
Favareto
V
.
Clinical practices in neonatal oxygenation: where have we failed? What can we do?
J Perinatol
.
2008
;
28
(
suppl 1
):
S28
S34
[PubMed]
36
Askie
LM
,
Henderson-Smart
DJ
,
Ko
H
.
Restricted versus liberal oxygen exposure for preventing morbidity and mortality in preterm or low birth weight infants.
Cochrane Database Syst Rev
.
2009
;(
1
):
CD001077
[PubMed]
37
Tin
W
,
Milligan
DW
,
Pennefather
P
,
Hey
E
.
Pulse oximetry, severe retinopathy, and outcome at one year in babies of less than 28 weeks gestation.
Arch Dis Child Fetal Neonatal Ed
.
2001
;
84
(
2
):
F106
F110
[PubMed]
38
Payne
NR
,
LaCorte
M
,
Karna
P
, et al;
Breathsavers Group, Vermont Oxford Network Neonatal Intensive Care Quality Improvement Collaborative
.
Reduction of bronchopulmonary dysplasia after participation in the Breathsavers Group of the Vermont Oxford Network Neonatal Intensive Care Quality Improvement Collaborative.
Pediatrics
.
2006
;
118
(
suppl 2
):
S73
S77
[PubMed]
39
The STOP-ROP Multicenter Study Group
.
Supplemental Therapeutic Oxygen for Prethreshold Retinopathy of Prematurity (STOP-ROP), a randomized, controlled trial. I: primary outcomes.
Pediatrics
.
2000
;
105
(
2
):
295
310
[PubMed]
40
Askie
LM
,
Henderson-Smart
DJ
,
Irwig
L
,
Simpson
JM
.
Oxygen-saturation targets and outcomes in extremely preterm infants.
N Engl J Med
.
2003
;
349
(
10
):
959
967
[PubMed]
41
Cole
CH
,
Wright
KW
,
Tarnow-Mordi
W
,
Phelps
DL
;
Pulse Oximetry Saturation Trial for Prevention of Retinopathy of Prematurity Planning Study Group
.
Resolving our uncertainty about oxygen therapy.
Pediatrics
.
2003
;
112
(
6 pt 1
):
1415
1419
[PubMed]
42
Askie
LM
,
Brocklehurst
P
,
Darlow
BA
,
Finer
N
,
Schmidt
B
,
Tarnow-Mordi
W
;
NeOProM Collaborative Group
.
NeOProM: Neonatal Oxygenation Prospective Meta-analysis Collaboration study protocol.
BMC Pediatr
.
2011
;
11
(
6
):
6
[PubMed]
43
Carlo
WA
,
Finer
NN
,
Walsh
MC
, et al;
SUPPORT Study Group of the Eunice Kennedy Shriver NICHD Neonatal Research Network
.
Target ranges of oxygen saturation in extremely preterm infants.
N Engl J Med
.
2010
;
362
(
21
):
1959
1969
[PubMed]
44
Vaucher
YE
,
Peralta-Carcelen
M
,
Finer
NN
, et al;
SUPPORT Study Group of the Eunice Kennedy Shriver NICHD Neonatal Research Network
.
Neurodevelopmental outcomes in the early CPAP and pulse oximetry trial.
N Engl J Med
.
2012
;
367
(
26
):
2495
2504
[PubMed]
45
Stenson
BJ
,
Tarnow-Mordi
WO
,
Darlow
BA
, et al;
BOOST II United Kingdom Collaborative Group
;
BOOST II Australia Collaborative Group
;
BOOST II New Zealand Collaborative Group
.
Oxygen saturation and outcomes in preterm infants.
N Engl J Med
.
2013
;
368
(
22
):
2094
2104
[PubMed]
46
Tarnow-Mordi
W
,
Stenson
B
,
Kirby
A
, et al;
BOOST-II Australia and United Kingdom Collaborative Groups
.
Outcomes of two trials of oxygen-saturation targets in preterm infants.
N Engl J Med
.
2016
;
374
(
8
):
749
760
[PubMed]
47
Cummings
JJ
,
Lakshminrusimha
S
,
Polin
RA
.
The BOOST trials and the pitfalls of post hoc analyses
.
N Engl J Med
.
2016
,
In press
48
Darlow
BA
,
Marschner
SL
,
Donoghoe
M
, et al
Randomized controlled trial of oxygen saturation targets in very preterm infants: two year outcomes. J Pediatr.
2014
;165(1):30–35.e2
49
Schmidt
B
,
Whyte
RK
,
Asztalos
EV
, et al;
Canadian Oxygen Trial (COT) Group
.
Effects of targeting higher vs lower arterial oxygen saturations on death or disability in extremely preterm infants: a randomized clinical trial.
JAMA
.
2013
;
309
(
20
):
2111
2120
[PubMed]
50
Stenson
B
,
Brocklehurst
P
,
Tarnow-Mordi
W
;
UK BOOST II trial
;
Australian BOOST II trial
;
New Zealand BOOST II trial
.
Increased 36-week survival with high oxygen saturation target in extremely preterm infants.
N Engl J Med
.
2011
;
364
(
17
):
1680
1682
[PubMed]
51
Saugstad
OD
,
Aune
D
.
Optimal oxygenation of extremely low birth weight infants: a meta-analysis and systematic review of the oxygen saturation target studies.
Neonatology
.
2014
;
105
(
1
):
55
63
[PubMed]
52
Manja
V
,
Lakshminrusimha
S
,
Cook
DJ
.
Oxygen saturation target range for extremely preterm infants: a systematic review and meta-analysis.
JAMA Pediatr
.
2015
;
169
(
4
):
332
340
[PubMed]
53
Guyatt
GH
,
Oxman
AD
,
Schünemann
HJ
,
Tugwell
P
,
Knottnerus
A
.
GRADE guidelines: a new series of articles in the Journal of Clinical Epidemiology.
J Clin Epidemiol
.
2011
;
64
(
4
):
380
382
[PubMed]
54
Walsh
MC
,
Di Fiore
JM
,
Martin
RJ
,
Gantz
M
,
Carlo
WA
,
Finer
N
.
Association of oxygen target and growth status with increased mortality in small for gestational age infants: further analysis of the Surfactant, Positive Pressure and Pulse Oximetry Randomized Trial
.
JAMA Pediatr
.
2016
;
170
(
3
):
292
294
[PubMed]
55
Bateman
D
,
Polin
RA
.
A lower oxygen-saturation target decreases retinopathy of prematurity but increases mortality in premature infants.
J Pediatr
.
2013
;
163
(
5
):
1528
1529
[PubMed]
56
Thomas
WJ
,
Rauser
M
,
Dovich
JA
,
Dustin
L
,
Flaxel
CJ
.
Oxygen saturation in premature infants at risk for threshold retinopathy of prematurity.
Eur J Ophthalmol
.
2011
;
21
(
2
):
189
193
[PubMed]
57
Di Fiore
JM
,
Bloom
JN
,
Orge
F
, et al
.
A higher incidence of intermittent hypoxemic episodes is associated with severe retinopathy of prematurity.
J Pediatr
.
2010
;
157
(
1
):
69
73
[PubMed]
58
Kaufman
DA
,
Zanelli
SA
,
Gurka
MJ
,
Davis
M
,
Richards
CP
,
Walsh
BK
.
Time outside targeted oxygen saturation range and retinopathy of prematurity.
Early Hum Dev
.
2014
;
90
(
suppl 2
):
S35
S40
[PubMed]
59
York
JR
,
Landers
S
,
Kirby
RS
,
Arbogast
PG
,
Penn
JS
.
Arterial oxygen fluctuation and retinopathy of prematurity in very-low-birth-weight infants.
J Perinatol
.
2004
;
24
(
2
):
82
87
[PubMed]

James J. Cummings, MD, FAAP

Richard A. Polin, MD, FAAP

Kristi L. Watterberg, MD, FAAP, Chairperson

Brenda Poindexter, MD, FAAP

James J. Cummings, MD, FAAP

William E. Benitz, MD, FAAP

Eric C. Eichenwald, MD, FAAP

Brenda B. Poindexter, MD, FAAP

Dan L. Stewart, MD, FAAP

Susan W. Aucott, MD, FAAP

Jay P. Goldsmith, MD, FAAP

Karen M. Puopolo, MD, PhD, FAAP

Kasper S. Wang, MD, FAAP

Richard A. Polin, MD, FAAP

Waldemar A. Carlo, MD, FAAP

Waldemar A. Carlo, MD, FAAP

Tonse N.K. Raju, MD, DCH, FAAP – National Institutes of Health

CAPT Wanda D. Barfield, MD, MPH, FAAP – Centers for Disease Control and Prevention

Erin L. Keels, APRN, MS, NNP-BC – National Association of Neonatal Nurses

Thierry Lacaze, MD – Canadian Paediatric Society

James Goldberg, MD – American College of Obstetricians and Gynecologists

Jim R. Couto, MA

Competing Interests

FINANCIAL DISCLOSURE: The authors have indicated they do not have a financial relationship relevant to this article to disclose.